FDA Recall Terminated

DeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, general hospital

Recall: Z-0098-2015 · Initiated August 22, 2014

Recall

Recall Number
Z-0098-2015
Event Number
69319
Firm
DeRoyal Industries Inc
FEI Number
1043214
Product Code
KDQ
Status
Terminated
Root Cause
Process control
Initiated
August 22, 2014
Posted
October 20, 2014
Terminated
April 15, 2015
Address
200 Debusk Ln, Powell, TN, 37849-4703

Description

DeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, general hospital

Reason

These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition.

Action

DeRoyal sent an Urgent Voluntary Recall letter dated August 22, 2014, to all affected consignees by overnight letter or by email. DeRoyal was to follow up with non-responding accounts before October 1, 2014. Customers were instructed to check their inventory for the affected product. If they have affected product, customers were instructed to destory both the liner and the lid and dispose of it according to their facility's procedures. Customers were instructed to return the included Notice of Destruction to [email protected] even if they no longer have affected product. Customers with questions were instructed to contact their DeRoyal Sales Representative or DeRoyal Customer Service at 1-800-251-9864. For questions regarding this recall call 865-362-2334.

Distribution

Nationwide Distribution including MI, OH, IN, FL, WV, VA, MI, NC, and NY

Quantity

5112 units