smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
Recall: Z-0097-2018
·
Initiated October 24, 2017
Recall
- Recall Number
- Z-0097-2018
- Event Number
- 78390
- Firm
- Smith & Nephew, Inc.
- FEI Number
- 1020279
- Product Code
- JDS
- Status
- Open, Classified
- Root Cause
- Employee error
- Initiated
- October 24, 2017
- Address
- 1450 E Brooks Rd, Memphis, TN, 38116-1804
Description
smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
Reason
A single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down.
Action
Smith & Nephew sent an Urgent Medical Device Recall Notice dated October 24, 2017, to all affected customers. The firm requested return of the product. Please follow the instructions on the attached Response Form. For further questions, please call (978) 749-1440.
Distribution
US Distribution - TN, IL, CT, MO, CA., and Internationally to Italy and Germany
Quantity
17 units