PIPELINE LS ACCESS SYSTEM, Holding Clip The PIPELINE LS Dilators and Holding Clip are intended for sequential dilation of soft tissue and enabling proper placement of the Pipeline LS Retractor.
Recall
- Recall Number
- Z-0097-2014
- Event Number
- 65897
- Firm
- DePuy Spine, Inc.
- FEI Number
- 1526439
- Product Code
- GCC
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- July 31, 2013
- Posted
- October 23, 2013
- Terminated
- June 29, 2015
- Address
- 325 Paramount Dr, Raynham, MA, 02767-5199
Description
PIPELINE LS ACCESS SYSTEM, Holding Clip The PIPELINE LS Dilators and Holding Clip are intended for sequential dilation of soft tissue and enabling proper placement of the Pipeline LS Retractor.
Dilators and Holding Clip are being used for a secondary function of neuromonitoring, for which it is not 510(k) cleared.
DePuy Synthes sent an Urgent Voluntary Product Recall Notification letter dated July 30, 2013, to all affected consignees via Fed Ex overnight delivery. Consignees were informed that DePuy Spine, Inc. is voluntarily recalling all lots of the PIPELINE LS ACCESS SYSTEM Dilators (1 st, 2nd, 3rd, and 4th) and Holding Clip. They were asked to review inventory located at their facility and quarantine any products listed in this recall notice. Consignees were instructed to contact their local DePuy Synthes Spine Sales Consultant to return any inventory located within their facility. Consignees with questions were instructed to call 508-828-3583.
Nationwide Distribution including the states of AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WV, and WY.
198 units