FDA Recall Terminated

Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for the reverse shoulder prosthesis head (glenosphere).

Recall: Z-0093-2021 · Initiated September 2, 2020

Recall

Recall Number
Z-0093-2021
Event Number
86422
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
PHX
Status
Terminated
Root Cause
Packaging process control
Initiated
September 2, 2020
Terminated
December 31, 2020
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for the reverse shoulder prosthesis head (glenosphere).

Reason

Product potentially being packaged without a taper adapter or baseplate. The package should contain one (1) baseplate and one (1) taper adapter, product may be packaged with two (2) baseplates and no (0) adapters or with no (0) baseplates and two (2) adapters

Action

Zimmer Biomet issued Urgent Medical Device Recall Letter dated 9/02/20 to Distributors via email and Hospital Risk Mangers via courier stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have unconsumed affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].

Distribution

US Nationwide distribution including in the states of CO, MN, OH, WY.

Quantity

3 units