FDA Recall Open, Classified

Servo-i Ventilator, Part No. 6487800, UDI Code 07325710000823 - Product Usage: intended for general and critical ventilator care, and it can be used for low end to high end intensive care, in different configurations. It supports a variety of breathing modes and can be used for Neonates, Pediatric and Adults.

Recall: Z-0092-2021 · Initiated September 17, 2020

Recall

Recall Number
Z-0092-2021
Event Number
86548
Firm
Getinge Group Logistics America, LLC
FEI Number
3025504317
Product Code
CBK
Status
Open, Classified
Root Cause
Device Design
Initiated
September 17, 2020
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Servo-i Ventilator, Part No. 6487800, UDI Code 07325710000823 - Product Usage: intended for general and critical ventilator care, and it can be used for low end to high end intensive care, in different configurations. It supports a variety of breathing modes and can be used for Neonates, Pediatric and Adults.

Reason

A potentially shorter than specified nebulizer connector may result in difficulty in installing the Aeroneb Nebulizer module to the Servo-i device successfully.

Action

On September __, 2020, the firm issued an Urgent Medical Device Recall letter to affected customers. Distributors were instructed to forward the information to all current and potential Servo-i Ventilator System users and biomedical staff within hospitals/facilities. Customers were asked to examine their inventory immediately to determine if they have any products with the listed serial numbers. For affected units, an alternate means of nebulization would need to be readily available and used until the unit correction can be performed by the Getinge Service Technician if nebulization via the Servo-i Aeroneb Nebulization module be required. Customers were also asked to complete and return the response form to the firm to acknowledge their receipt of the notification. If you have one or more affected systems, you will be contacted by a Getinge Service Representative to schedule on-site correction of your device as soon as an update kit is available. Affected devices will be corrected free of charge. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9438872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). You may also call this number if you have any questions about this recall.

Distribution

Domestic distribution to AZ CA GA IL MN MO NC NJ NY RI SC TX VA WI WV Foreign distributed to Bosnia and Herzegovina, Czech Republic, France, Germany, India, Indonesia, Iraq, Italy, Kenya, Republic of Korea, Lebanon, Libyan Arab Jamahiriya, Norway, Palestine, Peru, Qatar, Saudi Arabia,Slovakia, Sri Lanka, Sweden Thailand, Turkmenistan, Ukraine and United Arab Emirates.

Quantity

431 units