Wallach LL100 Cryosurgical Freezer CO2 System Model: 9000002X.
Recall
- Recall Number
- Z-0092-2010
- Event Number
- 53436
- Firm
- Cooper Surgical, Inc.
- FEI Number
- 1216677
- Product Code
- GEH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 29, 2009
- Posted
- October 29, 2009
- Terminated
- April 17, 2012
- Address
- 95 Corporate Dr, Trumbull, CT, 06611
Description
Wallach LL100 Cryosurgical Freezer CO2 System Model: 9000002X.
CO2 Cryosystem thermal load may exceed the exhaust gas phase, resulting in ice formation in the exhaust valve. Should this occur, the Cryo tip may cool at a less than optimal rate, resulting in ice formation in the exhaust valve.
Cooper Surgical notified customers by letter titled "Notice of Recall - Urgent" dated September 29, 2009 advising them of the problem and requesting return of the device. International accounts were notified by letter Beginning October 1, 2009. For further information, contact Cooper Surgical by fax at 1-203-799-2002 or phone at 1-203-799-2000.
Worldwide Distribution -- United States, Canada, Australia, Belgium, England, France, Germany, Greece, Hungary, Iceland, Italy, Ireland, Latvia, Lithuania, Netherlands, Portugal, Spain, Solvenia, Sweden, Hong Kong, Republic of China, Taiwan, Angola, Cameroon, Congo, Croatia, Egypt, Ghana, Guam, Guineae Bissau, Indonesia, Kenya, Kuwait, Lebanon, Madagascar, Mali, Mauritania, Mozambique, New Zealand, Niger, Pakistan,Qatar, Russia, Sao Tome, Saudi Arabia, Senegal, Serbia, Singapore, South Africa, St Vincents & Grenadines, Oman, Thailand, Turkey, UAE and Zimbabwe.