SoftPath GUI version 3.17.6.3.1. SCC Soft Computer 34350 US Highway 19 N, Palm Harbor, FL 34684. Date of Manufacture: Version 3.17.6.3.1 06/03/2003.
Recall
- Recall Number
- Z-0090-2011
- Event Number
- 56280
- Firm
- SCC Soft Computer
- FEI Number
- 1000307651
- Product Code
- LNX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 11, 2004
- Posted
- October 19, 2010
- Terminated
- November 3, 2010
- Address
- 5400 Tech Data Drive, Clearwater, FL, 33760
Description
SoftPath GUI version 3.17.6.3.1. SCC Soft Computer 34350 US Highway 19 N, Palm Harbor, FL 34684. Date of Manufacture: Version 3.17.6.3.1 06/03/2003.
For clients using the SoftPath GUI version 3.17.6.3.1 Results Reporting interface with parsed text and having the auto correct and auto formatting functions active in MS Word, there is a possibility for some ANSI characters/symbols to be translated incorrectly to ASCII. SCC Soft Computer initiated a Correction in March 2004 for all affected clients.
SCC Soft Computer issued a Risk-to-Health notification with Client Notification Verbiage to the affected client using SCC Soft Computer's proprietary Task Management System on March 11, 2007. SCC recommended turning off the autocorrect and autoformatting functions in MS Word if custom configuration parses text to an ASCII file (HIS) for the Result Reporting interface. If this function was turned on at the time of creating the canned message in the setup file, edit and re-save all canned messages in SoftPath. Client was asked to acknowledge receipt of the task and grant permission to load the required software correction.
Nationwide distribution: One hospital in NY.
1