Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-209-500); the bottles are sold alone and with various cap & float assemblies: 0321-7986-800 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0018-800 - Bottle with V Wall Bracket, Cap & Float; 6700-0370-800 - Bottle with Floor Bracket, Cap & Float. The collection bottles are part of a suction bottle assembly which is designed to collect various body fluids.
Recall
- Recall Number
- Z-0090-2010
- Event Number
- 53192
- Firm
- Ohio Medical Corporation
- FEI Number
- 1419185
- Product Code
- KDQ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 21, 2009
- Posted
- October 29, 2009
- Terminated
- June 22, 2011
- Address
- 1111 Lakeside Dr, Gurnee, IL, 60031
Description
Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-209-500); the bottles are sold alone and with various cap & float assemblies: 0321-7986-800 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0018-800 - Bottle with V Wall Bracket, Cap & Float; 6700-0370-800 - Bottle with Floor Bracket, Cap & Float. The collection bottles are part of a suction bottle assembly which is designed to collect various body fluids.
The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.
Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.
Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, Philippines, Qatar, Saudi Arabia, Singapore, Thailand, Turkey, United Kingdom and Venezuela.
823 bottles