FDA Recall Terminated

Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.

Recall: Z-0089-06 · Initiated October 7, 2005

Recall

Recall Number
Z-0089-06
Event Number
33816
Firm
Medrad Inc
FEI Number
2520313
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
October 7, 2005
Posted
October 25, 2005
Terminated
November 29, 2007
Address
1 Medrad Dr, Indianola, PA, 15051-9759

Description

Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.

Reason

overhead system can fall due to the failure of a retaining ring.

Action

The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/7/05 to all customers. The notice informs customers of the problem and cautions anyone to be under the system when in use. The notice also informs the customer that a firm representative will schedule a time to perform the upgrade. On 10/13/05, the recalling firm also issued a notification to scanner manufacturers alerting them of the problem.

Distribution

The product was shipped to medical facilities nationwide. The product was also shipped to Austria, Belgium, Chile, Czech, Denmark, Spain, Finland, France, Great Britain, Ireland, Israel, Italy, Netherlands, Norway, Sweden, Austrailia, Kuwait, Saudia Arabia, Canada, China, Hong Kong, India, Korea, Singapore, Taiwan, Brazil, Columbia, Cuba, Mexico, and Japan.

Quantity

2080 units