FDA Recall Terminated

SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured 03/18/2010. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.

Recall: Z-0088-2011 · Initiated July 19, 2010

Recall

Recall Number
Z-0088-2011
Event Number
56851
Firm
SCC Soft Computer
FEI Number
1000307651
Product Code
LNX
Status
Terminated
Root Cause
Software design
Initiated
July 19, 2010
Posted
October 19, 2010
Terminated
July 16, 2012
Address
5400 Tech Data Drive, Clearwater, FL, 33760

Description

SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured 03/18/2010. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.

Reason

Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities. Incomplete culture results may appear complete. SCC Soft Computer is issuing a Correction for SoftMic Versions 4.0.4.7-4.5.1.3 software due to this potential hazard.

Action

SCC Soft Computer issued Risk-to-Health notifications to clients on July 19, 2010 via proprietary communication software. Corrective actions will be via programming correction via hot fix (software modification) or such so that Global variable is not used. Clients are asked to acknowledge receipt of this task and update the task with permission to load the required correction.

Distribution

Worldwide Distribution: USA, including the states of CA, CT, DE, FL, IA, KY, MA, MA, MI, MN, NJ, NY, OH, PA, TX, VA, and WI, and the country of Canada.

Quantity

35