FDA Recall
Terminated
Y-valve and T-valve, PSD/4 Syringe Drive Module parts.
Recall: Z-0087-06
·
Initiated August 24, 2005
Recall
- Recall Number
- Z-0087-06
- Event Number
- 33435
- Firm
- Hamilton Co
- FEI Number
- 2915796
- Product Code
- JJC
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 24, 2005
- Posted
- October 22, 2005
- Terminated
- March 24, 2006
- Address
- 4970 Energy Way, Reno, NV, 89502-4123
Description
Y-valve and T-valve, PSD/4 Syringe Drive Module parts.
Reason
The product valve has a high probability of deforming and may result in leakage during use.
Action
Recall initiated on 08/24/2005. The firm issued notification to its consignees via overnight letter, and via e-mail for its distributor (Hamilton Bonaduz).
Distribution
One distributor and 8 consumers received this product in CA, MA, MI, WI, Japan and Switzerland.
Quantity
271 Y and 78 T valves were distributed.