FDA Recall Terminated

RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649

Recall: Z-0087-04 · Initiated October 6, 2003

Recall

Recall Number
Z-0087-04
Event Number
27500
Firm
Rita Medical Systems
FEI Number
1000523114
Product Code
GEI
Status
Terminated
Root Cause
Other
Initiated
October 6, 2003
Posted
November 1, 2003
Terminated
May 5, 2004
Address
967 N Shoreline Blvd, Mountain View, CA, 94043-1932

Description

RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649

Reason

The product displays inaccurate temperature readings, patient skin burns can occur before getting to higher temperature reading.

Action

On 10/6/03, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected devices and providing instructions on the recall.

Distribution

Product was released for distribution to 66 consignees in US. The product received nationwide distribution. The recall was appropriately extended to the user level; i.e., the hospitals that received the recalled product. There were no shipments to foreign or Canadian distribution, however, it includes one military distribution: Wilford Hall Medical Center, 2200 Berquist Dr., 31, Lackland AFB, TX 78236.

Quantity

137 + 225 units (lot Y082003)