FDA Recall Open, Classified

Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples

Recall: Z-0086-2025 · Initiated September 4, 2024

Recall

Recall Number
Z-0086-2025
Event Number
95323
Firm
LEICA BIOSYSTEMS NUSSLOCH GMBH Heidelberger Str. 17-19 Nussloch Germany
FEI Number
3002179321
Product Code
IDP
Status
Open, Classified
Root Cause
Labeling design
Initiated
September 4, 2024
Posted
October 15, 2024

Description

Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples

Reason

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Action

Leica Biosystems issued an Urgent Field Action Notice to its consignees on 09/05/2024 via mail. The notice explained the problem with the device and requested the placement of a warning sticker regarding the use of flammable freezing sprays on the device and the addition of a supplemental sheet to the current IFU. For questions, please contact: Heather Jensen, Leica Biosystems, 5205 US Highway 12, Richmond, IL 60071 USA, Tel.: +1-815-298-8371, [email protected]

Distribution

US Nationwide distribution.

Quantity

13 units