FDA Recall Open, Classified

cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190

Recall: Z-0083-2022 · Initiated September 3, 2021

Recall

Recall Number
Z-0083-2022
Event Number
88607
Firm
Roche Molecular Systems, Inc.
FEI Number
2243471
Product Code
OWD
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
September 3, 2021
Address
1080 US Highway 202 S, Branchburg, NJ, 08876-3733

Description

cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190

Reason

False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment

Action

Roche issued Urgent Medical Device Correction letter on 9/3/21 via UPS Ground. Letter states reason for recall, health risk and action to take: Follow the cobas DNA Sample Preparation Kit (catalog number 05985536190) IFU for sample input with regard to the number of FFPET sections and DNA concentration. Follow the cobas cfDNA Sample Preparation Kit (catalog number 07247737190) IFU for plasma samples. " If an Ex20Ins Mutation Detected result is generated with the cobas EGFR Mutation Test v2, you must confirm the result with another method (e.g., sequencing or other PCR-based tests). " Consider reviewing results generated since May 2021. Amivantamab was approved by the U.S. Food and Drug Administration (FDA) for NSCLC patients with EGFR Ex20Ins on May 21, 2021. Test result reports with Ex20Ins Mutation Detected may be considered for confirmatory testing using sequencing or other PCR-based tests, upon the discretion of a CAP/CLIA laboratory director. " Consult with your facility physician and/or pathologist to determine any clinical implications specific to your patients. " Complete all sections of the enclosed faxback form (TP-01381) and fax it to 1-855-208-4038 or email it to [email protected]. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-1247 if you have questions about the information contained in this UMDC

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Canada, UK, Poland, Spain, Italy, Sweden, Greece, Denmark, Norway, Netherlands, France, Czech Republic, Belgium, Austria, Slovakia, Hungary, Switzerland, Lithuania, Slovenia. Serbia, North Macedonia, Myanmar, Lebanon, Russia, Croatia, Morocco, Kazakhstan, Dominican Republic, Israel, El Salvador, Vietnam, Malaysia, Taiwan, China, Japan, Hong Kong, Korea, Brazil, Philippines, Thailand, Chile, Colombia, Singapore, Argentina, Australia, India, Uruguay, Ecuador, Panama, Costa Rica, Pakistan, Peru, Indonesia, New Zealand.

Quantity

507 US; 6,676 ex- US