Instrument Manager -Software Versions 8.05, 8.06, 8.07, 8.08, and 8.09. It is a stand-alone software package that provides connectivity between laboratory instruments and Laboratory Information Systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information. Instrument Manager also provides data management functions that aid in the interpretation and management of laboratory results. The Instrument Manager software is put on a CD which is distributed in a box. User documentation consists of user manuals, PDF copies of which are on the CD Intended to be used by trained laboratory personnel as an aid in laboratory instrument interfacing and laboratory data management.
Recall
- Recall Number
- Z-0083-2012
- Event Number
- 52105
- Firm
- Data Innovations, Inc.
- FEI Number
- 3001451449
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 11, 2009
- Posted
- October 21, 2011
- Terminated
- October 25, 2011
- Address
- 120 Kimball Ave, Ste 100, South Burlington, VT, 05403-6837
Description
Instrument Manager -Software Versions 8.05, 8.06, 8.07, 8.08, and 8.09. It is a stand-alone software package that provides connectivity between laboratory instruments and Laboratory Information Systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information. Instrument Manager also provides data management functions that aid in the interpretation and management of laboratory results. The Instrument Manager software is put on a CD which is distributed in a box. User documentation consists of user manuals, PDF copies of which are on the CD Intended to be used by trained laboratory personnel as an aid in laboratory instrument interfacing and laboratory data management.
The "previous result" retrieved from the Specimen Management Database may not be the most recent "previous result".
Data Innovations sent an "IMPORTANT PRODUCT NOTICE" dated May 11, 2009 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter offers temporary workarounds for customers to use until corrected versions of the software is complete. Versions of the corrections will be provided on a staggered schedule. Customers will be notified by e-mail. Questions regarding this notice are directed to Support at 802-658-1955.
USA (nationwide) and the country of Canada.