FDA Recall Terminated

BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen Calibrator Sets, REF: 663-1101

Recall: Z-0082-2022 · Initiated November 1, 2019

Recall

Recall Number
Z-0082-2022
Event Number
88652
Firm
Bio-Rad Laboratories, Inc.
FEI Number
3022521
Product Code
LKJ
Status
Terminated
Root Cause
Process control
Initiated
November 1, 2019
Terminated
September 27, 2023
Address
6565 185th Ave Ne, Redmond, WA, 98052-5039

Description

BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen Calibrator Sets, REF: 663-1101

Reason

A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.

Action

On 11/01/2019, ANA Screen Lot Matching-Compatibility Notice, were mailed to customers. Customers were advised of the following: If you have received the new control material, please ensure that your laboratory is using one of the newer lot numbers of BioPlex ANA calibrators and ANA reagents. Should you have any of the older BioPlex ANA reagent packs and calibrator sets listed below in ITALICS remaining in your inventory or if you have used these lots in combination with the new control, please contact your local Technical Support center. The following telephone number was provided: 510-724-7000.

Distribution

Worldwide distribution - U.S Nationwide distribution including in the states of CA, FL, NJ, IL, NE, TN, IN, TX, LA, WA, MI, MA, MN, OH, HI, MO, PA, CT, UT, SD, NY, AL, PR, OR, MD, KY, NC, AZ, SC, VA, WI, GA, IA, RI, AK, OK, KS, ND, MS, ME, NM, AR and the countries of ES, AE, DE, FR, KR, CN, TW, JP, IL, SG, AU, NZ, HK, IT, CA, PT, GB, NO, CZ, SE, CH, UA, RU, BE, DK.

Quantity

206