External power supply adaptors that accompany Clinitek Status Connect System The Clinitek Status system is a urine analyzer.
Recall
- Recall Number
- Z-0082-2015
- Event Number
- 69228
- Firm
- Siemens Healthcare Diagnostics, Inc
- FEI Number
- 1219913
- Product Code
- JIL
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 19, 2014
- Posted
- October 17, 2014
- Terminated
- July 20, 2015
- Address
- 333 Coney St, East Walpole, MA, 02032-1516
Description
External power supply adaptors that accompany Clinitek Status Connect System The Clinitek Status system is a urine analyzer.
The external power supply for the Clintek Status analyzer, which is provided separately, is damaged. It can result in an electric shock to the User.
Siemens sent an Urgent Field Safety Notice dated September 2014, to all by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on September 19, 2014 both in the United States and elsewhere for communication with affected customers. This notice informs customers of the issue with the power supply and the potentially affected products that may contain a damaged power supply. Consignees are asked to complete a Field Correction Effectiveness Check form and return in to Siemens Healthcare Diagnostics via Fax to (312) 275-7795. All affected domestic customers were sent a hard copy of the Urgent Field Safety Notice via Federal Express on 9/23/14. All affected customers outside the US will be provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. Customers with questions were instructed to contact their local Siemens technical support representative. For questions regarding this recall call 877-229-3711.
Worldwide Distribution - USA (nationwide) including Puerto Rico and Internationally Canada, Mexico, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Great Britain, Greece, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Latvia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Sweden, Switzerland, Tadjikistan, Taiwan, Turkey, Turkmenistan, Unit.Arab Emir., United Kingdom, Vietnam.
approx. 4200