Leveen Coaccess Needle Electrode (3.5/15) and (4.0/15). Disposable, monopolar, electrosurgical devices, intended for use in conjunction with BSC radiofrequency generator for the thermal coagulation necrosis of soft tissues, including partial or complete ablation of nonresectable liver lesions.
Recall
- Recall Number
- Z-0080-2013
- Event Number
- 63336
- Firm
- Boston Scientific Corporation
- FEI Number
- 3005099803
- Product Code
- JOS
- Status
- Terminated
- Root Cause
- Labeling Change Control
- Initiated
- September 17, 2012
- Posted
- October 18, 2012
- Terminated
- July 17, 2013
- Address
- 100 Boston Scientific Way, Marlborough, MA, 01752-1234
Description
Leveen Coaccess Needle Electrode (3.5/15) and (4.0/15). Disposable, monopolar, electrosurgical devices, intended for use in conjunction with BSC radiofrequency generator for the thermal coagulation necrosis of soft tissues, including partial or complete ablation of nonresectable liver lesions.
Leveen Standard and Leveen Coaccess Needle Electrode System, Directions for Use(DFU) in Portuguese, are incorrect. all other languages in the DFU are correct. Consignees in Brazil, Portugal,and Hong-Kong are being notified.
Boston Scientific sent notifications, dated September 17, 2012, to each account in Brazil, Portugal and Hong Kong affected by the recall. The affected accounts were provided the corrected pages of the Portuguese DFU as part of the communication to ensure that those users that rely on the Portuguese DFU have the correct information.
Distributed in Brazil, Portugal and Hong Kong.
19,716 units distributed total