Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.
Recall: Z-0079-2008
·
Initiated March 30, 2007
Recall
- Recall Number
- Z-0079-2008
- Event Number
- 38311
- Firm
- Intuitive Surgical, Inc.
- FEI Number
- 3001675293
- Product Code
- NAY
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 30, 2007
- Posted
- February 21, 2008
- Terminated
- September 5, 2008
- Address
- 950 Kifer Rd, Sunnyvale, CA, 94086-5206
Description
Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.
Reason
System Lock-Up: Software anomalies could cause product failure during use, or on start-up. System transitions to a safe "soft-lock" state.
Action
Intuitive Field Service Engineers installed a software upgrade beginning on 03/30/2007, and left a notification letter that the upgrade had been performed at each site. New software level is A5.1 P2.
Distribution
Worldwide Distribution - USA, Belgium, China, France, India, Italy, Japan, Korea, Spain, Sweden, Switzerland, UK and Mexico.
Quantity
159