FDA Recall Open, Classified

Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.

Recall: Z-0078-2023 · Initiated September 16, 2022

Recall

Recall Number
Z-0078-2023
Event Number
90876
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
HHK
Status
Open, Classified
Root Cause
Labeling design
Initiated
September 16, 2022
Posted
October 12, 2022
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.

Reason

Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with the VC-10, which can lead to serious injuries as it does not fit properly with the VC-10.

Action

On September 16, 2022, the recalling firm notified affected customers through "URGENT Medical Device Field Corrective Action" letters. Customers were informed that the following new warning will be added to the IFU: Do not use replacement parts or accessories that are not compatible with the VC-10 system. Use of noncompatible accessories could lead to reduced system functionality and/or patient injury such as perforation, bleeding, and may require additional intervention. Refer to Section 11 for all accessories compatible with the VC-10 system. Customers should notify their users about risks associated with using non-compatible collection canisters and/or accessories and remind them to follow the list of compatible accessories outlined in Section 11 of the IFU. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. " Go to https://olympusamerica.com/recall. Enter the recall number 0414 " Enter your name and mailing address. Olympus will supply a copy of the revised IFU at a later date If you have distributed these devices outside your facility, please notify your customers of this matter immediately. If you have any questions, call (647) 999-3203 or email: [email protected] for any additional information or support concerning this matter.

Distribution

Domestic distribution nationwide. Foreign distribution to Canada and Japan.

Quantity

4184 US; 191 OUS