Medtronic Disposable Pressure Display Set, Models 61000, 61006, 62000, 62006. Sterilized by Ethylene Oxide. Non-pyrogenic. Coiled PVC tubing with threaded transducer dome (acrylic) with silicone diaphragm on proximal end. Product utilizes a stopcock on proximal and/or distal end. Distal end of tubing features a male luer port. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.
Recall
- Recall Number
- Z-0078-2012
- Event Number
- 59948
- Firm
- Medtronic Perfusion Systems
- FEI Number
- 1000116158
- Product Code
- DXS
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 9, 2011
- Posted
- October 20, 2011
- Terminated
- October 11, 2012
- Address
- 7611 Northland Dr N, Brooklyn Park, MN, 55428-1088
Description
Medtronic Disposable Pressure Display Set, Models 61000, 61006, 62000, 62006. Sterilized by Ethylene Oxide. Non-pyrogenic. Coiled PVC tubing with threaded transducer dome (acrylic) with silicone diaphragm on proximal end. Product utilizes a stopcock on proximal and/or distal end. Distal end of tubing features a male luer port. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.
Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.
193,590 units