FDA Recall Open, Classified

LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC

Recall: Z-0071-2024 · Initiated August 23, 2023

Recall

Recall Number
Z-0071-2024
Event Number
92999
Firm
DKK Dai-Ichi Shomei Co., Ltd. 1 Chome 32-26 Sakashita Itabashi Japan
FEI Number
3006437518
Product Code
FSY
Status
Open, Classified
Root Cause
Device Design
Initiated
August 23, 2023
Posted
October 6, 2023

Description

LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC

Reason

The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.

Action

An URGENT RECALL notification letter dated 8/23/23 was sent to the customer. Actions to be Taken by the Manufacturer We will provide an additional metal cover for the LFS bracket so that even if the shaft shears off, it will not pop out from the cover as well as instructions to properly install the cover. Also, the additional metal cover provides additional support to the LFS bracket and prevents the stopper block from broken or missing even if the excessive force is added to the bracket. The physical correction will be announced later. Other Information Contact information for questions: TEL: +81-3-3969-9221 FAX: +81-3-3969-9269 Email: [email protected] Monday through Friday, 8:00 AM to 5:00 PM, Japan Standard Time.

Distribution

US States: MI

Quantity

256 units