FDA Recall Terminated

Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by strategically placed mounting arms. Additionally, there are three port supports the user incorporates to position the lines that enter and exit the venous bag. The Backplate/Holder is designed to allow the user to move the port support location to accommodate different tubing circuit designs to permit mounting flexibility.

Recall: Z-0071-2013 · Initiated October 2, 2009

Recall

Recall Number
Z-0071-2013
Event Number
63225
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
KRI
Status
Terminated
Root Cause
Labeling design
Initiated
October 2, 2009
Posted
October 16, 2012
Terminated
January 29, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by strategically placed mounting arms. Additionally, there are three port supports the user incorporates to position the lines that enter and exit the venous bag. The Backplate/Holder is designed to allow the user to move the port support location to accommodate different tubing circuit designs to permit mounting flexibility.

Reason

Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one inch connector supports instead of one 3/8 inch and two inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a

Action

The firm, Terumo, interviewed the Sr. Product Manager from the Marketing Department of the company on August 29, 2012. The interview included the description of the product, problem and action taken. Replacement connectors were provided to users and all units were corrected. If you have any questions, call 734-663-4145 or Manager, Regulatory Affairs at 734-741-5576.

Distribution

Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.

Quantity

33 units