FDA Recall
Open, Classified
GS0008 2Y - Burn Pack - STL, Item Number 880082025
Recall: Z-0070-2024
·
Initiated September 11, 2023
Recall
- Recall Number
- Z-0070-2024
- Event Number
- 93077
- Firm
- ROi CPS LLC
- FEI Number
- 3014527682
- Product Code
- OJJ
- Status
- Open, Classified
- Root Cause
- Other
- Initiated
- September 11, 2023
- Posted
- October 6, 2023
- Address
- 3000 E Sawyer Rd, Republic, MO, 65738-2758
Description
GS0008 2Y - Burn Pack - STL, Item Number 880082025
Reason
Medical convenience kit contains Zimmer Dermatome Blades which were recalled by Zimmer due to 38 complaints received related to skin grafts being thin and non-uniform when using the affected blades.
Action
ROi CPS, LLC issued a Medical Device Correction notice to its sole consignee on 09/11/2023 by email. The notice explained the issue with the Zimmer Dermatone Blade contained in the kit, the risk, and requested the consignee affix a sticker to the exterior of the kit requesting that the blade be removed and discarded at the time of use.
Distribution
Missouri
Quantity
103 kits