FDA Recall Terminated

Gemini GXL PET/CT Systems, 6, 10, 16 slice configurations. (Product Numbers/manufacturer numbers: 16 slice: 4535 679 71891, 10 slice: 4535 679 75691, 6 slice: 4535 679 75691. Catalog numbers: 16 slice: 882410, 10 slice: 882400, 6 slice: 882390.)

Recall: Z-0068-06 · Initiated August 18, 2005

Recall

Recall Number
Z-0068-06
Event Number
33273
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
August 18, 2005
Posted
October 21, 2005
Terminated
October 29, 2008
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

Gemini GXL PET/CT Systems, 6, 10, 16 slice configurations. (Product Numbers/manufacturer numbers: 16 slice: 4535 679 71891, 10 slice: 4535 679 75691, 6 slice: 4535 679 75691. Catalog numbers: 16 slice: 882410, 10 slice: 882400, 6 slice: 882390.)

Reason

The image balance center point correction algorithm used in CT image reconstruction may cause a disapperance or alteration of small circular anatomical structures that are both near the image center and perpendicular to the image plane. This artifact may lead to the misinterpretation of anatomy at the center of the image.

Action

The recalling firm sent a recall/field correction letter, dated August 18, 2005 to customers, notifying them of the problem.

Distribution

The product was distributed to customers within the US and to foreign customers located in Canada, France, Thailand, South Africa, and Italy.

Quantity

19 devices