INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language variant. IN = English (International); GB=English (US); FR= French; DE=German; IT=Italian; PT=Portuguese; ES=Spanish; JP=Japanese; PL=Polish; RU=Russian
Recall
- Recall Number
- Z-0062-06
- Event Number
- 33276
- Firm
- Event Medical
- FEI Number
- 3006135941
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 12, 2005
- Posted
- October 15, 2005
- Terminated
- February 29, 2012
- Address
- 2440 Grand Ave, Vista, CA, 92081-7829
Description
INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language variant. IN = English (International); GB=English (US); FR= French; DE=German; IT=Italian; PT=Portuguese; ES=Spanish; JP=Japanese; PL=Polish; RU=Russian
Firm received reports of combustion within the blender system of the Inspiration Ventilator.
Customer shall be notified by an Urgent Product Safety Notfication letter dated August 12, 2005. The letter shall advise the customer with the specifics of this problem, summarise the fact that the hazard potential in negligible and will advise the customer regarding options for replacement of the Blender Module assembly (part number F710400) and the Air/O2 Filters (part number F910300). The letter will also provide the customers with a set of actions which can be completed immediately in order to ensure the integrity of their gas supplies until their unit has been upgraded.
Nationwide and Canada, Europe, Japan, Latin America, Asia/Pacific, Middle East
1,602 units