FDA Recall Terminated

Symbia S, Product Usage 1) SPECT: To Detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. 2) CT: The CT component is intended to produce corss-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. 3) SPECT + CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patients anatomy. .

Recall: Z-0061-2017 · Initiated August 16, 2013

Recall

Recall Number
Z-0061-2017
Event Number
66544
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
1000514697
Product Code
KPS
Status
Terminated
Root Cause
Component design/selection
Initiated
August 16, 2013
Terminated
December 23, 2016
Address
2501 Barrington Rd, Hoffman Estates, IL, 60195-2061

Description

Symbia S, Product Usage 1) SPECT: To Detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. 2) CT: The CT component is intended to produce corss-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. 3) SPECT + CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patients anatomy. .

Reason

The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.

Action

Customers were contacted by a service organization that performed product inspections at their sites.

Distribution

Worldwide Distribution - US: Nationwide, including Washington D.C., and Puerto Rico. and the countries of: Poland, France, Austria, Czech Republic, Ireland, Germany, Italy, Japan, Canada, United Arab Emirates, Turkey, Netherlands, United Kingdom, Brazil, South Africa, Kuwait, Belgium, Russia, Taiwan, Australia, Bangladesh, Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, Hong Kong, Croatia, Hungary, Indonesia, Iran, Israel, India, South Korea, Kuwait, Kazakhstan, Lithuania, Latvia, Morocco, Mexico, Norway, New Zealand, Peru, Phillipines, Pakistan, Portugal, Qatar, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Taiwan, Venezuela, and Vietnam.

Quantity

732 units