Medical Gas Yoke Assembly. Airgas Portable 2-Cylinder Manifold Cart Model #AGI-2237. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy.
Recall
- Recall Number
- Z-0061-2010
- Event Number
- 52847
- Firm
- Western a Scott Fetzer Company
- FEI Number
- 1526809
- Product Code
- CAN
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- August 10, 2009
- Posted
- October 15, 2009
- Terminated
- April 25, 2012
- Address
- 875 Bassett Rd, Westlake, OH, 44145-1142
Description
Medical Gas Yoke Assembly. Airgas Portable 2-Cylinder Manifold Cart Model #AGI-2237. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy.
The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.
Western Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.
Nationwide Distribution -- AL, AR, CA, CO, FL, GA, HI, IL, IN, MA, OH, OR, PA, TX, WA and WI.
42 cart assemblies containing 84 yoke assemblies (42 Oxygen yokes and 42 Nitrous Oxide yokes)