Stat Profile Critical Care Expres Ionized Magnesium Sensor (In-vitro Diagnostic used with the Stat Profile Critical Care Xpress Analyzer) Part # 37271
Recall
- Recall Number
- Z-0061-04
- Event Number
- 27334
- Firm
- Nova Biomedical Corporation
- FEI Number
- 1219029
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 22, 2003
- Posted
- October 24, 2003
- Terminated
- February 4, 2004
- Address
- 200 Prospect Street, Waltham, MA, 02454
Description
Stat Profile Critical Care Expres Ionized Magnesium Sensor (In-vitro Diagnostic used with the Stat Profile Critical Care Xpress Analyzer) Part # 37271
Degradation of the magnesium sensor could result in low ionozed magnesium patient samples
Nova Biomedical contacted customers by telephone beginning on 9/26/03. Each account was faxed or mailed Customer Information Bulletin requesting the discard of Mg++ sensors in inventory that are greater than 4 months old from date of manufacture. Replacement product will be issued.
Natinwide Foreign: Italy, Czech, Japan, Taiwan, Hungary, India, Israel, Mexico, Canada, Germany,Switzerland, Tunisia, Korea, Brazil, Greece, Coatia, China, France, Turkey, Thailand, Argentina. Government: VA Decator, GA
280 units