FDA Recall Open, Classified

DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.

Recall: Z-0060-2021 · Initiated August 3, 2020

Recall

Recall Number
Z-0060-2021
Event Number
86257
FEI Number
1000116912
Product Code
MBH
Status
Open, Classified
Root Cause
Process change control
Initiated
August 3, 2020
Posted
October 7, 2020
Address
9800 Metric Blvd, Encore Medical, LP, Austin, TX, 78758-5445

Description

DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.

Reason

There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.

Action

On August 3, 2020, Field Safety Notices were sent to the firm's distributors by email advising them of this field action, informing them that the firm has received two product complaints where an issue has been experienced with the retaining pin becoming loose and falling out while impacting. Distributors were informed that no adverse events have been reported but that there is a potential of leaving debris in the joint space if this issue were to occur and remain undetected. The letter requests distributors to communicate the following information to implanting surgeons to avoid the potential of leaving debris in the joint space: 1. Sales representative is to inspect instrumentation before every surgery per the Instructions For Use. A small object should be pushed against the end of the pin to confirm it is securely fixed in place. 2. The sales rep should warn of the possibility of the loose pin and steps to take to confirm containment of the loose pin in order to mitigate patient risk. 3. After use of the instrument, it should be verified that the pin remains fixed in place. Distributors were informed that when devices that meet specification become available for the 5 impacted lots, a field exchange will occur as part of a reportable recall. Distributors are requested to complete the acknowledgment form and return it by 8-10-2020 to confirm that they have read and understood best practices for use of the knee punch handle, and have disseminated the information to all employees and contracted Sales Representatives. Customers with questions should contact Teffany Hutto by phone at (512) 834-6255 or by email at [email protected]. On 03/12/2021, recall notices were emailed to additional customers informing them of additional affected models/lots.

Distribution

U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.

Quantity

190 units