Philips Medical System - Endura X-Ray Systems (Release 2) with Remote Control Option
Recall: Z-0060-2008
·
Initiated August 23, 2007
Recall
- Recall Number
- Z-0060-2008
- Event Number
- 39392
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- JAA
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 23, 2007
- Posted
- November 10, 2007
- Terminated
- September 27, 2010
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
Philips Medical System - Endura X-Ray Systems (Release 2) with Remote Control Option
Reason
System Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.
Action
On 8/23/07, the firm sent out a Safety Notification letter, via certified mail. On 8/29/07, the firm sent out an URGENT - DEVICE CORRECTION letter via regular mail, dated 8/20/07. Both letters advised customers that if the device shuts down during use, it can be restarted within about 1 minute by switching the system off and then on again, using the system on/off buttons. The letter(s) states that software is being developed to correct the problem.
Distribution
Nationwide
Quantity
37 units