FDA Recall Terminated

Unicel Closed Tube Aliquotter (UCTA) for UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, UniCel DxC 880i Synchron Access Clinical Systems. Part Number A45853. Product Usage: The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in-vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).

Recall: Z-0059-2013 · Initiated July 6, 2012

Recall

Recall Number
Z-0059-2013
Event Number
63191
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 6, 2012
Posted
October 12, 2012
Terminated
November 15, 2012
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Unicel Closed Tube Aliquotter (UCTA) for UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, UniCel DxC 880i Synchron Access Clinical Systems. Part Number A45853. Product Usage: The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in-vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).

Reason

The recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause a hardware motion error (ID0x08000501).

Action

The recall notification was initiated on July 6, 2012 with Beckman Coulter forwarding a Product Corrective Action (PCA) letter with attached PCA Response Form to all customers who purchased the UniCel DxC 880i, UniCel DxC 860i, UniCel 680i, UniCel DxC 660i Clinical System Integrated Workstations UCTA Sample Syringe Pump Drive Assembly. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were notified that Beckman Coulter will schedule a Service Call to replace the affected sample syringe pump drive assemblies. Customers were instructed to complete and return the enclosed PCA Response Form within 10 days. Customers with any questions regarding the PCA, were instructed to call Customer Support at (800) 854-3633 in the US and Canada. Outside the US and Canada, customers were instructed to contact Beckman Coulter Representative.

Distribution

Worldwide Distribution and US (nationwide) and the countries of China, France, Italy and Taiwan.

Quantity

18 units