FDA Recall Terminated

Medical Gas Yoke Assembly. Generic 4-Cylinder Manifold Cart Model #8070. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy.

Recall: Z-0059-2010 · Initiated August 10, 2009

Recall

Recall Number
Z-0059-2010
Event Number
52847
Firm
Western a Scott Fetzer Company
FEI Number
1526809
Product Code
CAN
Status
Terminated
Root Cause
Component design/selection
Initiated
August 10, 2009
Posted
October 15, 2009
Terminated
April 25, 2012
Address
875 Bassett Rd, Westlake, OH, 44145-1142

Description

Medical Gas Yoke Assembly. Generic 4-Cylinder Manifold Cart Model #8070. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy.

Reason

The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.

Action

Western Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.

Distribution

Nationwide Distribution -- AL, AR, CA, CO, FL, GA, HI, IL, IN, MA, OH, OR, PA, TX, WA and WI.

Quantity

18 cart assemblies containing 72 yoke assemblies (36 Oxygen yokes and 36 Nitrous Oxide yokes)