FDA Recall Terminated

Philips Medical System - BV Pulsera Mobile C-arms with Remote Control Option, (Release 2)

Recall: Z-0059-2008 · Initiated August 23, 2007

Recall

Recall Number
Z-0059-2008
Event Number
39392
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
JAA
Status
Terminated
Root Cause
Software design
Initiated
August 23, 2007
Posted
November 10, 2007
Terminated
September 27, 2010
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Philips Medical System - BV Pulsera Mobile C-arms with Remote Control Option, (Release 2)

Reason

System Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.

Action

On 8/23/07, the firm sent out a Safety Notification letter, via certified mail. On 8/29/07, the firm sent out an URGENT - DEVICE CORRECTION letter via regular mail, dated 8/20/07. Both letters advised customers that if the device shuts down during use, it can be restarted within about 1 minute by switching the system off and then on again, using the system on/off buttons. The letter(s) states that software is being developed to correct the problem.

Distribution

Nationwide

Quantity

596 units