BD Microtainer SST- Amber Part/Catalog Number: 365978
Recall
- Recall Number
- Z-0058-2020
- Event Number
- 82851
- Firm
- Becton Dickinson & Company
- FEI Number
- 2243072
- Product Code
- JKA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- May 13, 2019
- Terminated
- August 6, 2020
- Address
- 1 Becton Dr, Franklin Lakes, NJ, 07417-1815
Description
BD Microtainer SST- Amber Part/Catalog Number: 365978
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
Becton Dickinson (BD) notified consignees affected by letter via UPS and email dated May 13,2019. Letter describes reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific catalog and lot numbers listed below, 2. Destroy all product subject to the recall following your institution's process for destruction, 3. Share this Urgent Medical Device Recall notification with all users of the product to ensure that they are also aware of this recall, and 4. Complete the attached Customer Response/Certificate of Destruction Forms and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Distributors provide their customer list to BD. BD PAS will notify these customers via mail. BD Contact US Contact Information Customer/Technical Support 888-237-2762 OPT 3, OPT 2; Monday Friday 8:00am - 5:00pm (CT).
Nationwide Foreign: AR AU BE BN BR CA CL CN CO CR CW EC GT GY HK HN ID IN JP KR MX MY NI NZ PA PE PG PH PK SG SR SV TH TW UY VN