THERMOCOOL SMARTTOUCH Uni-Directional Navigation Catheter (D133602S)

Recall: Z-0057-2018 · Initiated September 6, 2017

Recall

Recall Number
Z-0057-2018
Event Number
78129
Firm
Biosense Webster, Inc.
FEI Number
3003113629
Product Code
LPB
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
September 6, 2017
Terminated
November 19, 2019
Address
15715 Arrow Hwy, Irwindale, CA, 91706-2006

Description

THERMOCOOL SMARTTOUCH Uni-Directional Navigation Catheter (D133602S)

Reason

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.

Action

All US and OUS customers, who have purchased the lots identified as part of this recall will be notified via customer letters in order to return the product to BWI facility where it will be segregated for further disposition. This product recall applies to hospitals (or direct customers) and distributors who have purchased and/or received the THERMOCOOL Catheters identified as part of this recall . One notification package which includes the Customer Notification Letter, Business Reply Form (B RF) and pre-paid return label will be distributed to the customers subject to this recall via a 3rd party vendor (Stericycle) for USA. OUS customer communication will occur in accordance with local regulations.

Distribution

US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China

Quantity

19 units