Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO.

Recall: Z-0056-2008 · Initiated August 24, 2007

Recall

Recall Number
Z-0056-2008
Event Number
39408
Firm
Mallinckrodt Inc
FEI Number
1518293
Product Code
DXT
Status
Terminated
Root Cause
Device Design
Initiated
August 24, 2007
Posted
October 13, 2007
Terminated
August 17, 2010
Address
2111 E Galbraith Rd, Cincinnati, OH, 45237-1624

Description

Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO.

Reason

Synchronization problems: units that have both a main programmable console and an auxiliary programmable console, when simultaneously, injection parameters are programmed into the unit's main console and the unit's auxiliary consoles, mis-synchronization could occur resulting in a variance between the intended injection parameters (those programmed) and the actual parameters of the injection deliv

Action

On 9/20/2007, a notification was sent from Mallinckrodt to all direct accounts via certified mail. The notification details the potential for mis-synchronization of injection parameters if the main and auxiliary consoles are simultaneously programmed. The notification reminds end-users that simultaneous programming should not be performed and to ensure it is not performed, a notification label (provided in the notification) is to be affixed on both consoles of the injector system. Once the label is affixed to the system, end-users are instructed to complete the notification survey and return the survey to the Mallinckrodt. In addition, the notification includes an updated chapter insert (Chapter 11.1) for the injector system's product manual that emphasize the separation between programming of the main and auxiliary consoles. Questions regarding this matter may be directed to Mallinckrodt's Product Monitoring Department at 1-888-744-1414, menu option 1 and then menu option 2.

Distribution

Worldwide to include:U.S.A. states of - AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, & WI. and Foreign countries of Argentina, Belgium, Brazil, Canada, Germany, Japan, Singapore, & Zambia.

Quantity

176