3 PORT "OFF" MANIFOLD w/WASTE BAG, SYRINGE and DyePod, Part Number (Item #) AG8049-NS, Lot # 2528655. K914299. Waste system.
Recall
- Recall Number
- Z-0055-2013
- Event Number
- 63362
- Firm
- ICU Medical, Inc.
- FEI Number
- 2025816
- Product Code
- DQO
- Status
- Terminated
- Root Cause
- Mixed-up of materials/components
- Initiated
- September 28, 2012
- Posted
- October 11, 2012
- Terminated
- February 19, 2013
- Address
- 951 Calle, Amanecer San Clemente, CA, 92673-6212
Description
3 PORT "OFF" MANIFOLD w/WASTE BAG, SYRINGE and DyePod, Part Number (Item #) AG8049-NS, Lot # 2528655. K914299. Waste system.
ICU Medical, Inc. has identified a potential comingling of components where a 60 drop micro drip chamber may have been included on a 20 drop macro drip set.
ICU Medical sent an "URGENT:MEDICAL DEVICE RECALL" letter dated September 28, 2012 to its customer. The letter identified the product, problem, and actions to be taken by the customer. The customer has been instructed to inspect their inventory for the specific lot numbers, examine for the presence of a metal cannula, and certify destruction of any affected devices. The customer letter included a Verification Form to be completed and returned. Contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time: 866-829-9025 and select option 8, or e-mail the following address: [email protected]" for questions regarding this notice.
Nationwide distribution: Virginia only.