FDA Recall Terminated

ApexPro FH Telemetry System: composed of six major components (as follows); Accessories to the patient worn acquisition transceivers, the patient worn data acquisition transceivers, the transceiver access points with antenna, the network infrastructure, A computer platform running the ApexPro Telemetry Application and a computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application). GE Medical Systems information Technologies, 8200 West Tower Avenue, Milwaukee, WI 53223.

Recall: Z-0054-2007 · Initiated June 23, 2006

Recall

Recall Number
Z-0054-2007
Event Number
35826
Firm
Ge Healthcare
FEI Number
3005860720
Product Code
MHX
Status
Terminated
Root Cause
Other
Initiated
June 23, 2006
Posted
October 21, 2006
Terminated
May 25, 2007
Address
9900 W Innovation Dr, Wauwatosa, WI, 53226-4856

Description

ApexPro FH Telemetry System: composed of six major components (as follows); Accessories to the patient worn acquisition transceivers, the patient worn data acquisition transceivers, the transceiver access points with antenna, the network infrastructure, A computer platform running the ApexPro Telemetry Application and a computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application). GE Medical Systems information Technologies, 8200 West Tower Avenue, Milwaukee, WI 53223.

Reason

System Warning Alarm failure: When a patient being monitored is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC does not occur.

Action

An Urgent Medical Device Correction letter, dated June 23, 2006, was sent to customers notifying them of the issue, provided short term recommendations and provided the long term solution as a software upgrade that will correct this situation. A reply card is asked to be returned to initiate the upgrade process.

Distribution

Nationwide distribution --- including states of AL, AR, AZ, CA, CO, Washington DC, FL, IL, IN, ME, MI, MN, MO, NC, NE, NJ, NY, OK, PA, SC, TX, VA, WA, WI, WV.

Quantity

379