FDA Recall Terminated

PermaFlo Flowable Composite REF-UP 948 Shade: A2 - Product Usage: is a light cured, radiopaque, methacrylate based, flowable composite. PermaFlo contains H1m average particle size with narrow upper limit particle distribution.

Recall: Z-0053-2021 · Initiated August 31, 2020

Recall

Recall Number
Z-0053-2021
Event Number
86432
Firm
Ultradent Products, Inc.
FEI Number
1718912
Product Code
EBF
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
August 31, 2020
Terminated
October 18, 2021
Address
505 W 10200 S, South Jordan, UT, 84095-3800

Description

PermaFlo Flowable Composite REF-UP 948 Shade: A2 - Product Usage: is a light cured, radiopaque, methacrylate based, flowable composite. PermaFlo contains H1m average particle size with narrow upper limit particle distribution.

Reason

Due to a potential manufacturing issue (cross contamination), composite is non-homogeneous.

Action

On August 31, 2020, the firm sent an URGENT FIRST NOTICE PRODUCT RECALL" notification to US customers via UPS (International customer via email on September 1, 2020) informing them of the potential product contamination that may affect the performance characteristics of the product to include, micro-hardness, bond strength, and wear and may pose an increased risk of restoration failure. Customers are instructed to discontinue use of the product immediately. Customers are also instructed to: -Complete the Return Response Form enclosed with this letter and Fax 801-553-4609 or email to [email protected]. -Using the supplied shipping labels, return to Ultradent the completed form and all unused Product; and -Maintain a ledger of the patients treated with the Product, including the patients name and applicable treatment information. The Recalling Firm will send out replacement kits and issue an $100 credit to customer's accounts For questions regarding replacement product, contact Customer Service at 1-888-230-1420. For any questions regarding the recall, contact Customer Concerns at 801-553-4139.

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, KS, IN, LA, MA, MD, ME, MI, MT, NC, NJ, NM, NY, NV, OH, OK, PA, SC, TN, TX, UT, VT, WA, and WI. The countries of Australia, Germany Japan and Mongolia.

Quantity

412 syringes (206 kits)