THERMI250 Radiofrequency System. Intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions.
Recall
- Recall Number
- Z-0052-2018
- Event Number
- 77754
- Firm
- Bios S.r.l. Via Guido Rossa 10/12 Vimodrone Italy
- FEI Number
- 3008108972
- Product Code
- PBX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 13, 2017
- Terminated
- September 15, 2020
Description
THERMI250 Radiofrequency System. Intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions.
In some systems, the unit continues to emit RF energy after the PAUSE button is pressed.
Consignees were sent a recall notification by mail dated 07/13/2017. Customers were notified of the reason for recall and advised that the PAUSE button not be used and the STOP button be used instead to stop the RF emission. Additionally, units are to be connected to the internet so that the BIOS service team can access the units to provide a software update to correct the issue. If a unit cannot be connected to the internet, customers are to contact Thermi to schedule an update.
AZ, AR, CA, FL, GA, ID, IL, KS, LA, MA, MI, MS, MO, NV, NJ, NM, NY, OH, OR, PA, SD, TX, WA, WI, and WV Brazil, Canada, Israel, Japan, & Thailand.
167 units