FDA Recall Open, Classified

StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures.

Recall: Z-0050-2025 · Initiated August 22, 2024

Recall

Recall Number
Z-0050-2025
Event Number
95283
Firm
Medtronic Navigation, Inc.
FEI Number
1000517638
Product Code
OLO
Status
Open, Classified
Root Cause
Software design
Initiated
August 22, 2024
Posted
October 10, 2024
Address
200 Medtronic Drive, Lafayette, CO, 80026

Description

StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures.

Reason

Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.

Action

On 08/22/2024, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was mailed to customers informing them of 2 software anomalies impacting spine procedures on the StealthStation S8 system (StealthStation S8 Basic, StealthStation S8 Plus and StealthStation S8 Premium) with Application software version 1.3.0 or 1.3.2. Issues are: 1.Navigation Orientation: When lateral orientation radiographic images are taken using a navigation-enabled Ziehm Vision FD Vario 3D C-arm, the orientation information appears flipped when displayed by the StealthStation S8 system. The navigation information is displayed with a 180-degree rotation around the superior/inferior(S/I) axis with the left/right (L/R) axis and anterior/posterior (A/P) axis appearing as mirror images. The exam image orientation and related indication labels on the exams remain correct. Note -this anomaly only impacts StealthStation" S8 systems where the Ziehm Vision FDVario 3D C-arm Interface Kit, 9734486, installed and calibrated on a Ziehm FD system in the lateral orientation; other orientations settings are not impacted. Medtronic Recommends Do not use the lateral orientation setting on the Ziehm Vision FD Vario 3D C-arm, if using the StealthStation" S8 with software version 1.3 or 1.3.2. 2.Pop-up Message: When using CT + Fluoroscopy images with a slice spacing thickness greater than 2 mm, the system does not display a pop-up message indicating the slice spacing is not optimal. Navigation remains accurate in this situation. Customers are instructed to: -Review Appendix A which provides detailed examples of the issues and how to identify if your system is affected. -Medtronic will work expeditiously with facilities to coordinate a software update to version 2.1 at no charge to resolve anomalies. Questions, contact Medtronic Technical Services at 1-888-826-5603 or email at [email protected].

Distribution

Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TN, TX, UT, VA, WA, WI and WV. The countries of Argentina, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, Brazil, BULGARIA, CANADA, CHILE, China, Colombia, Costa Rica, CROATIA, CYPRUS, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, ICELAND, India, ISRAEL, ITALY, Japan, Jordan, KAZAKHSTAN, KENYA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, Malaysia, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PANAMA, Philippines, POLAND, PUERTO RICO, QATAR, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, Singapore, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, Turkey, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM.

Quantity

1,801 systems