FDA Recall Terminated

Toshiba America Medical Systems, Inc . Celesteion PCA-9000A/2 CT Scanner

Recall: Z-0049-2017 · Initiated April 25, 2016

Recall

Recall Number
Z-0049-2017
Event Number
74871
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
KPS
Status
Terminated
Root Cause
Software design
Initiated
April 25, 2016
Posted
October 6, 2016
Terminated
July 13, 2017
Address
2441 Michelle Dr PO Box 2068, Tustin, CA, 92780-7047

Description

Toshiba America Medical Systems, Inc . Celesteion PCA-9000A/2 CT Scanner

Reason

It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not occurred in the U.S.

Action

Toshiba America Medical System, Inc. planned action to bring the defect into Compliance: 1. A notification will be sent to customers with affected systems in compliance with 21 CFR 1003.21 upon approval by the Secretary. The notification letter includes a statement by the correction will be made at no cost to the customer. 2. To correct this issue, modified software will be installed on the systems to prevent this occurrence from happening. The consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. 3. To avoid this problem in scan planning, the consignee will be advised to do the following until the modified software is installed: a. Confirm that the scan range does not extend beyond the maximum FOV. b. If the ROI for the scan range (indicated by the yellow solid lines) is set as the same size as the maximum FOV (indicated by green dashed lines), the ROI position must be changed. Instructions for changing ROI will be provided as attachments to the notification letter. 4. The modification to the software will be completed within 90 days from the date that the letters are submitted to Toshibas customers. 5. Progress letters will be provided to the Secretary, including the number of electronic products repaired. For further questions, please call (714) 730-5000.

Distribution

US Distribution to the states of: PA, IA, MA, SD, MI, LA, KY, NC, NY, AR, IN, WI, OH, TY, FL, WV, WA, CT, MS, AL, ND, CO, HI, NO, MN, ID, OH, GA, MT, LA, NM, VT, TN, CA, NH, NJ, NE, SC and KS.

Quantity

165