DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)

Recall: Z-0049-2007 · Initiated October 3, 2006

Recall

Recall Number
Z-0049-2007
Event Number
36478
Firm
Del Medical Systems Group
FEI Number
1418964
Product Code
IZL
Status
Terminated
Root Cause
Other
Initiated
October 3, 2006
Posted
October 19, 2006
Terminated
January 24, 2008
Address
11550 West King Street, Franklin Park, IL, 60131-1330

Description

DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)

Reason

The x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.

Action

Del Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.

Distribution

Nationwide and Thailand

Quantity

53 units