GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator

Recall: Z-0048-2025 · Initiated September 3, 2024

Recall

Recall Number
Z-0048-2025
Event Number
95282
Firm
DATEX--OHMEDA, INC.
FEI Number
3014220748
Product Code
FMZ
Status
Open, Classified
Root Cause
Use error
Initiated
September 3, 2024
Posted
October 17, 2024
Address
9900 W Innovation Dr, Wauwatosa, WI, 53226-4856

Description

GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator

Reason

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

Action

GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/03/2024 via Fed-Ex. The notice explained the problem with the device, safety issue, and requested the following: Download the Service Manual Addendum (5971733) from the GE HealthCare Customer Documentation Portal: https://www.gehealthcare.com/support/manuals. Inspect the heater door mounting screw on your device. If the screw is not appropriately tightened, complete steps 9 and 10 of Section 1.1 of the Service Manual Addendum (5971733).

Distribution

Worldwide distribution.