STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
Recall
- Recall Number
- Z-0048-2014
- Event Number
- 66304
- Firm
- Abbott Medical Optics, Inc.
- FEI Number
- 3006695864
- Product Code
- LZS
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- September 16, 2013
- Posted
- October 22, 2013
- Terminated
- November 12, 2014
- Address
- 510 Cottonwood Dr, Milpitas, CA, 95035-7403
Description
STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
The firm, Abbott Medical Optics, Inc., (AMO), sent an "ADVISORY NOTICE and a COVER LETTER FOR ADVISORY NOTICE" dated September 16, 2013 to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. The customers were instructed to complete and return the Star Notification Acknowledgement form via fax to 1-714-247-4530 within 3 business days or mail to Abbott Medical Optics, 510 Cottonwood Drive, MS 347118, Milpitas, CA 95035, and share this advisory notice with your staff. If you have any questions related to operation of STAR Excimer Laser System or documentation requirements associated with this letter, please call AMO Technical Support at 1-800-511-0911 and you will be directed to an AMO Representative.
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.
834 units in total