FDA Recall Terminated

RESOLVE Hemoglobin Test Kits packed and distributed under the Perkin Elmer Life and Analytical Sciences label, Kit Codes: FR-9120 - 120 tests per kit; FR-9400 - 360 tests per kit; and FR-9360 - 3600 tests per kit.

Recall: Z-0047-05 · Initiated September 14, 2004

Recall

Recall Number
Z-0047-05
Event Number
30017
Firm
PerkinElmer LAS Inc
FEI Number
1522510
Product Code
GKA
Status
Terminated
Root Cause
Other
Initiated
September 14, 2004
Posted
October 28, 2004
Terminated
October 29, 2008
Address
3985 Eastern Rd, Norton, OH, 44203-6215

Description

RESOLVE Hemoglobin Test Kits packed and distributed under the Perkin Elmer Life and Analytical Sciences label, Kit Codes: FR-9120 - 120 tests per kit; FR-9400 - 360 tests per kit; and FR-9360 - 3600 tests per kit.

Reason

The product insert for the test kits is incorrect with regard to the acceptable storage temperature for the Agarose IEF gel component. The kit and IEF gel labels indicate the correct storage temperature of 2-8 degrees C. The product insert incorrectly indicates that the product may be stored at 2-30 degrees C.

Action

The firm contacted their consignees by letter dated 9/14/04.

Distribution

Nationwide (The product was distributed to firms located in AR, AZ, CA, CO,FL, GA, HI, IA, IL, IN, KS, KY,LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV )and worldwide (to 12 foreign accounts).

Quantity

All lots of test kits manufactured by the firm which have an active shelf life are subject to recall.