Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Recall: Z-0046-2018
·
Initiated August 3, 2017
Recall
- Recall Number
- Z-0046-2018
- Event Number
- 77982
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 3, 2017
- Terminated
- September 27, 2023
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Reason
Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
Action
BEC will initiate a SW update- to fix the SW defect. This will be executed via MOD (field modification) performed by Field Service.
Distribution
China France Italy Kuwait Spain Turkey United States
Quantity
27 units