FDA Recall Terminated

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

Recall: Z-0046-2018 · Initiated August 3, 2017

Recall

Recall Number
Z-0046-2018
Event Number
77982
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
August 3, 2017
Terminated
September 27, 2023
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

Reason

Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.

Action

BEC will initiate a SW update- to fix the SW defect. This will be executed via MOD (field modification) performed by Field Service.

Distribution

China France Italy Kuwait Spain Turkey United States

Quantity

27 units