FDA Recall Terminated

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Recall: Z-0044-2014 · Initiated September 18, 2013

Recall

Recall Number
Z-0044-2014
Event Number
66317
Firm
Advanced Sterilization Products
FEI Number
3003662624
Product Code
MLR
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
September 18, 2013
Posted
October 22, 2013
Terminated
January 13, 2014
Address
33 Technology Dr, Irvine, CA, 92618-2346

Description

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Reason

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

Action

Advanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers were also asked to maintain a copy of the notice with the affected product. Customers with questions or suspected problems were instructed to contact Stericycle directly at 1-877-257-7119. For questions regarding this recall call 949-453-6400.

Distribution

Nationwide Distribution

Quantity

2,995 units