FDA Recall Open, Classified

One Step Ovulation Test REF 200-01

Recall: Z-0043-2024 · Initiated May 22, 2023

Recall

Recall Number
Z-0043-2024
Event Number
92659
Firm
Universal Meditech Inc.
FEI Number
3011908678
Product Code
NGE
Status
Open, Classified
Root Cause
Other
Initiated
May 22, 2023
Posted
October 23, 2023
Address
850 I St, Reedley, CA, 93654-2819

Description

One Step Ovulation Test REF 200-01

Reason

Universal Meditech Inc. was violatively distributing PrestiBio" Breastmilk Alcohol Test Strip, DiagnosUS SARS-CoV-2 Antibody (IgG/IgM) Test, HealthyWiser KetoFast" Ketone Test Strips and PrestiBio" Ketone Test Strips without marketing authorization. UMI is also recall their tests because they are going out of business and would not be able to continue fulfilling any post-market responsibilities of these distributed products. UMI claims to hold 510(k)s for the other devices, purchased the intellectual property in 2015 from the previous 510(k) holder, but FDA has been unable to verify this information. The devices may have unknown performance characteristics because the firm is unable to provide any documentation that the devices were manufactured in conformance with the Quality System regulation, including documentation (e.g., DHF) that the distributed devices had not been modified since original clearance in a way that could impact their safety and effectiveness, documentation of controlled storage temperature/humidity, and post-market surveillance documentation, such as complaints and adverse events.

Action

On 06/07/2023, the firm emailed customers to inform them of the intent of the firm to recall all products we have sold to them in the past two years; in that email, UMI has advised them to immediately stop further distribution of any product that was made by UMI. In the past few weeks, UMI has conducted a thorough investigation of their order information. We will send formal recall letters to them with all the batch information of all products they have ordered since March 2021. On 06/30/2023, the firm sent via US Certified Post, a hard copy letter of it first version of the recall notification. On 09/02/2023, the firm sent an updated "URGENT: MEDICAL DEVICE RECALL" letter via email to inform customers that Universal Meditech Inc. is going out of business hence would not be able to continue fulfilling any post-market responsibilities for these distributed products. Therefore, the firm decided to recall any products that have not yet been sold to end users. Customers are being advised to immediately stop distributing or using any of the abovementioned devices in your remaining stock. If you have already distributed the device to any other distributors or retailors, we kindly ask you to inform them of the product recall by forwarding this letter to them. Please kindly acknowledge receipt of this recall notification by filling out the attached form at the end of this letter and sending it back to us. For questions, contact Michael M. Lin, Legal Attorney at 702-871-9888 (9AM to 5PM, PST, MON to FRI) or email [email protected] Firm issued a Press Release - posting at www.universal-meditech.com Update Information (09/15/2023): On 09/17/2023, additional "URGENT: MEDICAL DEVICE RECALL" Letter were sent via email and FedEx, to inform customer of additional products added to the Recall and to indicated that products being recalled are for any products that have not been used and instructing customers to notify customers that they have further distributed the products

Distribution

U.S.: CA, FL, IL, LA, NJ, NY, TX and WY O.U.S.: Taiwan

Quantity

100,000 tests